Intraocular Lenses
| Model (REF) | 92S |
| Total Diameter | 15.0 mm |
Body Diameter | 6.0 mm |
| Effective Optical Diameter | 5.5 mm |
| Position | Capsular Bag |
| Haptic | C-Loop | 10° |
| Standard Diopter Range | 10.0 – 30.0 D (0.5 inc.) |
| On Request Diopter Range | -15.0 – 9.5 D (0.5 inc.) 30.5 – 40.0 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 119.7 | Scleral sutured = 120.6 |
| Theoretical ACD (optical) | 5.96 mm | Scleral sutured = 6.48 mm |
| Haigis | a0 = 1.40 | a1 = 0.40 | a2 = 0.10 |
| SRK/T | SRK II | 119.1 | 119.6 |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
| Media | Surgery Video |
| Instruction For Use | eIFU-System |
| Model (REF) | 46G |
| Total Diameter | 15.0 mm |
Body Diameter | 6.0 mm |
| Effective Optical Diameter | 6.0 mm |
| Position | Capsular Bag |
| Haptic | C-Loop | 10° |
| Standard Diopter Range | 10.0 – 26.0 D (0.5 inc.) |
| On Request Diopter Range | -5.0 – 9.5 D (0.5 inc.) 26.5 – 35.0 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 119.7 | Scleral sutured = 120.6 |
| Theoretical ACD (optical) | 5.96 mm | Scleral sutured = 6.48 mm |
| Haigis | a0 = 1.40 | a1 = 0.40 | a2 = 0.10 |
| SRK/T | SRK II | 119.1 | 119.6 |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | Ultra Violet and High Energy Visible Light Blocker |
| Refractive Index | 1.46 |
| Instruction For Use | eIFU-System |
| Model (REF) | 90L |
| Total Diameter | 15.0 mm |
Body Diameter | 6.5 mm |
| Effective Optical Diameter | 6.0 mm |
| Position | Capsular Bag |
| Haptic | C-Loop | 10° |
| Standard Diopter Range | 10.0 – 30.0 D (0.5 inc.) |
| On Request Diopter-Range | 8.5 – 9.5 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 119.7 | Scleral sutured = 120.6 |
| Theoretical ACD (optical) | 5.96 mm | Scleral sutured = 6.48 mm |
| Optimized-A-Con. (optical) | Scleral sutured = 118.1 |
| Optimized ACD (optical) | Scleral sutured = 5.02 mm |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
| Instruction For Use | eIFU-System |
| Model (REF) | 92G |
| Total Diameter | 15.0 mm |
Body Diameter | 6.5 mm |
| Effective Optical Diameter | 6.0 mm |
| Position | Capsular Bag |
| Haptic | C-Loop | 10° |
| Standard Diopter Range | 10.0 – 30.0 D (0.5 inc.) |
| On Request Diopter Range | 8.5 – 9.5 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 119.7 | Scleral sutured = 120.6 |
| Theoretical ACD (optical) | 5.96 mm | Scleral sutured = 6.48 mm |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
| Instruction For Use | eIFU-System |
| Model (REF) | 92B |
| Total Diameter | 15.0 mm |
Body Diameter | 7.0 mm |
| Effective Optical Diameter | 6.5 mm |
| Position | Capsular Bag |
| Haptic | C-Loop | 10° |
| Standard Diopter Range | 10.0 – 30.0 D (0.5 inc.) |
| On Request Diopter Range | -10.0 – 9.5 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 119.7 | Scleral sutured = 120.6 |
| Theoretical ACD (optical) | 5.96 mm | Scleral sutured = 6.48 mm |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
| Note | Cannot be implanted with an injector due to its optic size. |
| Instruction For Use | eIFU-System |
| Model (REF) | 92C |
| Total Diameter | 10.5 mm |
Body Diameter | 6.0 mm |
| Effective Optical Diameter | 5.5 mm |
| Position | Capsular Bag |
| Haptic | 4-Loop | 10° |
| Standard Diopter Range | 15.0 – 30.0 D (0.5 inc.) |
| On Request Diopter Range | 8.5 – 14.5 D (0.5 inc.) 30.5 – 35.0 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 119.1 |
| Theoretical ACD (optical) | 5.61 mm |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
| Instruction For Use | eIFU-System |
| Model REF | 97B | ||
| Total Diameter | 11.0 mm | 10.8 mm | 10.5 mm |
Body Diameter | 6.5 mm | 6.3 mm | 6.0 mm |
| Effective Optical Diameter | 6.5 mm | 5.8 mm | 5.5 mm |
| Optic | convexo-concave | biconvex | biconvex |
| Position | Capsular Bag | Capsular Bag | Capsular Bag |
| Haptic | 4-Loop | 0° | 4-Loop | 0° | 4-Loop | 0° |
| Standard Diopter Range | - | - | 15.0 – 30.0 D (0.5 inc.) |
| On Request Diopter Range | -5.0 – 4.5 D (0.5 inc.) | 5.0 – 14.5 D (0.5 inc.) | 30.5 – 35.0 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 118.8 | 118.8 | 118.8 |
| Theoretical ACD (optical) | 5.43 mm | 5.43 mm | 5.43 mm |
| Optimized A-Con. SRKT (optical)*3 | 118.8 | 118.8 | 118.4 |
| Material | Hydrophilic Acrylic | Hydrophilic Acrylic | Hydrophilic Acrylic |
| Water Content | 28.0 % | 28.0 % | 28.0 % |
| Filter | UV-Filter | UV-Filter | UV-Filter |
| Refractive Index | 1.46 | 1.46 | 1.46 |
| Instruction For Use | eIFU-System | eIFU-System | eIFU-System |
| REF | 90S |
|---|---|
| Total Diameter | 15.0 mm |
Body Diameter | 6.4 mm |
| Effective Optical Diameter | 5.0 – 4.7 mm |
| Position | Capsular Bag |
| Haptic | C-Loop | 0° |
| Standard Diopter Range | 10.0 – 30.0 D. (0.5 inc.) |
| On Request Diopter Range | 8.5 – 9.5 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 118.05 |
| Theoretical ACD (optical) | 4.97 |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
| Model (REF) | 89A |
| Indication | Adults / Pediatrics |
Total Major/Minor Axis Haptics | 7.5 / 6.0 mm |
| Effective Optical Diamter | 5.0 mm |
| Standard Diopter Range | 10.0 – 30.0 D (0.5 inc.) |
| On Request Diopter-Range | -13.0 – 9.5 D (0.5 inc.) 30.5 – 35.0 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 118.2 |
| Theoretical ACD (optical) | 5.08 mm |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
Recommended Caliper Ring*2 for ACCC*1 | |
| Note | Surgeons must partake in prerequiste course before implantation! |
| Security Advice | Due to the possibility of iris capture it is recommended to keep the pupil in myosis in the early postoperative period. |
| Media | Surgery Videos |
| Instruction For Use | eIFU-System |
| Model (REF) | 89D |
| Indication | For short eyes (< 18.0 mm) and/or small white to white < 10.0 mm |
Total Major/Minor Axis Haptics | 6.5 / 5.25 mm |
| Effective Optical Diamter | 4.5 mm |
| Standard Diopter Range | - |
| On Request Diopter Range | 8.5 – 35.0 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 118.2 |
| Theoretical ACD (optical) | 5.08 mm |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
Recommended Caliper Ring*2 for ACCC*1 | |
| Note | Surgeons must partake in prerequiste course before implantation! |
| Security Advice | Due to the possibility of iris capture it is recommended to keep the pupil in myosis in the early postoperative period. |
| Media | Surgery Videos |
| Instruction For Use | eIFU-System |
| Model (REF) | 89F |
| Indication | Combined cataract and vitrectomy procedures or in eyes with large pupils |
Total Major/Minor Axis Haptics | 8.5 / 7.0 mm 7.5 / 6.0 mm |
| Effective Optical Diamter | 5.0 mm |
| Standard Diopter Range | 10.0 – 30.0 D (0.5 inc.) |
| On Request Diopter Range | 8.5 – 9.5 D (0.5 inc.) |
| Theoretical A-Con. (optical) | 118.2 |
| Theoretical ACD (optical) | 5.08 mm |
| Material | Hydrophilic Acrylic |
| Water Content | 28.0 % |
| Filter | UV-Filter |
| Refractive Index | 1.46 |
Recommended Caliper Ring*2 for ACCC*1 | |
| Note | Surgeons must partake in prerequiste course before implantation! |
| Security Advice | Due to the possibility of iris capture it is recommended to keep the pupil in myosis in the early postoperative period. |
| Media | Surgery Videos |
| Instruction For Use | eIFU-System |
*1 Anterior Continuous Curvolinear Capsulorhexis
*2 Gauge for Capsulorhexis
*3 by B. Golenvaux, MD